Fluidized Zone   | Note: The gas is passed upward through the fluidized bed at a high enough velocity to lift the particles. However, particles may be swept away from the tube chamber entirely if the velocity was set too high. Depends on the particle size, you may want to adjust the gas in pressure and flow rate until particles are lifted and move turbulently inside of the tube. |
Quartz Frit | - Frits with forced convection design can trap the particles within the central heating area
- Frit: 15~40 microns porous, 22mm O.D x 4mm thick
 |
Optional Tube | - You may consider SS310 high-temperature stainless steel tube for processing high-pressure materials.
- Max. working temperature 800ºC
- Max. Pressure 3 bars (built-in over-pressure valve which can be automatically reset and reuseable)
- Please click the picture below to see the design:
 |
Furnace Casing | |
Power | 1200W |
Voltage | AC 110V or 220V Single Phase, 50/60 Hz |
Max. Temperature | 1100oC |
Continuous Temperature | 1000oC |
Max. Heating Rate | <= 20 oC /min |
Tube Size and Materials  | - One quartz tube is included. Main tube with bottom frit and enlarged space for particles processing
- Two Ends: 25ODx 22 ID (mm); Central: 50O.D. x 44 I.D. (mm), please click the picture to the left for details
- Central enlarged area ensures the high aerodynamic performance
|
Sealing Flanges  | - Bottom flange: Regular designed 1'' bottom flange with 1/4'' tube fitting for introducing primary gas
- Top Flange: Special designed 1'' top flange with backstreaming secondary (precursor) gas inlet and exhaust, please click the picture to the left for details
|
Options | - Optional gas mixing system is available for CVD process.
- Oxygen Monitoring: An Oxygen Sensor can be used to monitor the oxygen level of gas(es) used in CVD systems for preventing or reducing oxidation. Please click the picture below to learn more
- Ultrasonic Atomizer can be integrated into the system for delivering sub-micron droplets to the hot zone Please click the bottom right picture for details
 |
Heating Zone Length | 200mm (single zone) |
Constant Heating Zone | 100mm (+/-2oC) |
Temperature Controller  | - 30 programmable segments for precise control of heating rate, cooling rate, and dwell time.
- Built-in PID Auto-Tune function with overheating & broken thermocouple protection.
- Over-temperature protection and alarm allow for operation without attendant(s).
- +/- 1 ºC temperature accuracy.
- Eurotherm 3000 precision temperature controller is optional with +/- 0.1ºC accuracy
|
Heating Elements | Fe-Cr-Al Alloy doped by Mo |
Fitting Ports | 1/4'' Tube Fittings were installed on both sides of the flange for standard connection to the gas supplier |
Dimensions | 340mm×300mm×1000mm ( W x L x H) |
Net Weight | ~20 Kg (excluding mobile cart) |
Shipping Dimensions | 48" x 40" 35" (L x W x H) |
Shipping Weight | 150 lbs |
Warranty | One year limited warranty (Consumable parts such as heating elements, processing tubes, and o-rings are not covered by the warranty, please order the replacement at related products below.) |
Laptop, software & WiFi Control (Optional) | - Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial)
- Labview Based Temperature Control System (EQ-MTS01) enables the user to edit temperature profiles, manage heat-treatment recipe, record and plot data for MTI furnaces
- A wireless remote control provides up to 300 meters of operating range
- The above features are available upon request at an extra cost (up to $1,000)
- Note: The software is only compatible with MTI’s Yudian Temperature Controller

|
Compliance  | - CE Certified
- NRTL or CSA certification is available at an extra cost
|
Warning Click here to learn the installation of a gas regulator | - Tube furnaces with quartz tube is designed for using under low positive pressure < 0.2 bars / 3 psi / 0.02 Mpa
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- Vacuum limit definition for all quartz tube furnaces: * Vacuum pressures may only be safely used up to 1000°C
|
Operation Instructions video |
|